| Summary | Part name: | 1000/270 BARR ED123-1 | |||||||||
| Measurement: All parts are measured at 4 points. Once at Center and then at three points near the edge. In an f/8 beam at AOI=0 | |||||||||||
| Measured 77K shift for T%: | -0.0177 | ||||||||||
| Measured 77K shift for CWL: | 3.5 | ||||||||||
| Measured 77K shift for FWHM: | -1.1 | ||||||||||
| CWL Theoretical AOI shift 0º-5º: | 3.3 | ||||||||||
| FWHM Theoretical AOI shift 0º-5º: | -1.05 | ||||||||||
| (following table represents filter data at 77K & AOI = 5 degrees, ALL RAW DATA in worksheets attached)) | |||||||||||
| Position | center | edge1 | edge2 | edge3 | STATUS | Spec's | |||||
| T% | 96.93 | 96.78 | 96.84 | 96.44 | PASS | > 75% | |||||
| CWL(nm) | 990.6 | 990.9 | 991.4 | 991 | PASS | 1000+/-30nm | |||||
| FWHM(nm) | 276.15 | 276.75 | 275.8 | 277.05 | PASS | 270+/-30nm | |||||