| Summary | Part name: | 1973/6.6 BARR ED112-1 | |||||||||
| Measurement: All parts are measured at 4 points. Once at Center and then at three points near the edge. In an f/8 beam at AOI=0 | |||||||||||
| Measured 77K shift for T%: | -1.86346 | ||||||||||
| Measured 77K shift for CWL: | 0.4 | ||||||||||
| Measured 77K shift for FWHM: | 0.1 | ||||||||||
| Theoretical AOI shift 0º-5º: | 1.44 | ||||||||||
| (following table represents filter data at 77K & AOI = 5 degrees, ALL RAW DATA in worksheets attached)) | |||||||||||
| Position | center | edge1 | edge2 | edge3 | STATUS | Spec | |||||
| T% | 84.37 | 83.90 | 83.88 | 83.59 | PASS | > 70% | |||||
| CWL(nm) | 1972.8 | 1973.04 | 1973.1 | 1972.71 | PASS | 1973+/-0.3 | |||||
| FWHM(nm) | 6.64 | 6.63 | 6.62 | 6.62 | PASS | 6.6+/-0.6 | |||||