Summary |
|
Part
name: |
1973/6.6 BARR
ED112-1 |
|
|
|
|
|
Measurement: All parts are measured at 4 points. Once at Center and
then at three points near the edge. In an f/8 beam at AOI=0 |
|
|
|
|
|
|
|
|
|
|
|
|
Measured 77K shift for T%: |
-1.86346 |
|
|
|
|
|
|
|
|
Measured 77K shift for CWL: |
0.4 |
|
|
|
|
|
|
|
|
Measured 77K shift for FWHM: |
0.1 |
|
|
|
|
|
|
|
|
Theoretical
AOI shift 0º-5º: |
1.44 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
(following
table represents filter data at 77K & AOI = 5 degrees, ALL RAW DATA in
worksheets attached)) |
|
|
Position |
center |
edge1 |
edge2 |
edge3 |
STATUS |
Spec |
|
|
|
|
|
T% |
84.37 |
83.90 |
83.88 |
83.59 |
PASS |
> 70% |
|
|
|
|
|
CWL(nm) |
1972.8 |
1973.04 |
1973.1 |
1972.71 |
PASS |
1973+/-0.3 |
|
|
|
|
|
FWHM(nm) |
6.64 |
6.63 |
6.62 |
6.62 |
PASS |
6.6+/-0.6 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|