Summary |
|
Part
name: |
1965/6.5 BARR
ED115-1 |
|
|
|
|
|
Measurement: All parts are measured at 4 points. Once at Center and
then at three points near the edge. In an f/8 beam at AOI=0 |
|
|
|
|
|
|
|
|
|
|
|
|
Measured 77K shift for T%: |
-1.1096 |
|
|
|
|
|
|
|
|
Measured 77K shift for CWL: |
0.3 |
|
|
|
|
|
|
|
|
Measured 77K shift for FWHM: |
0 |
|
|
|
|
|
|
|
|
Theoretical
AOI shift 0º-5º: |
1.43 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
(following
table represents filter data at 77K & AOI = 5 degrees, ALL RAW DATA in
worksheets attached)) |
|
|
|
Position |
center |
edge1 |
edge2 |
edge3 |
STATUS |
SPEC |
|
|
|
|
|
T% |
90.38 |
90.42 |
90.22 |
90.35 |
PASS |
> 70% |
|
|
|
|
|
CWL(nm) |
1965.07 |
1964.97 |
1964.97 |
1964.87 |
PASS |
1965+/-0.3 |
|
|
|
|
|
FWHM(nm) |
6.3 |
6.3 |
6.3 |
6.2 |
PASS |
6.5+/-0.6 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|