| Summary | Part name: | 1965/6.5 BARR ED115-1 | |||||||||
| Measurement: All parts are measured at 4 points. Once at Center and then at three points near the edge. In an f/8 beam at AOI=0 | |||||||||||
| Measured 77K shift for T%: | -1.1096 | ||||||||||
| Measured 77K shift for CWL: | 0.3 | ||||||||||
| Measured 77K shift for FWHM: | 0 | ||||||||||
| Theoretical AOI shift 0º-5º: | 1.43 | ||||||||||
| (following table represents filter data at 77K & AOI = 5 degrees, ALL RAW DATA in worksheets attached)) | |||||||||||
| Position | center | edge1 | edge2 | edge3 | STATUS | SPEC | |||||
| T% | 90.38 | 90.42 | 90.22 | 90.35 | PASS | > 70% | |||||
| CWL(nm) | 1965.07 | 1964.97 | 1964.97 | 1964.87 | PASS | 1965+/-0.3 | |||||
| FWHM(nm) | 6.3 | 6.3 | 6.3 | 6.2 | PASS | 6.5+/-0.6 | |||||