H2r-2.13

Summary Part name:  2129.7/7 BARR ED089-1
Measurement: All parts are measured at 4 points. Once at Center and then at three points near the edge. In an f/8 beam at AOI=0
Measured 77K shift for T%:  -3.33175
Measured 77K shift for CWL:  0.3
Measured 77K shift for FWHM:  0
Theoretical AOI shift 0º-5º: 1.56
(following table represents filter data at 77K & AOI = 5 degrees, ALL RAW DATA in worksheets attached))
Position center edge1 edge2 edge3 STATUS
T% 76.61 76.57 76.31 76.59 PASS
CWL(nm) 2129.64 2129.64 2129.94 2130.04 PASS
FWHM(nm) 7.4 7.4 7.4 7.4 PASS