| Summary | Part name: | 2121.8/7 BARR ED086-1 | ||||||||||
| Measurement: All parts are measured at 4 points. Once at Center and then at three points near the edge. In an f/8 beam at AOI=0 | ||||||||||||
| Measured 77K shift for T%: | 0.418137 | |||||||||||
| Measured 77K shift for CWL: | 0.35 | |||||||||||
| Measured 77K shift for FWHM: | 0.1 | |||||||||||
| Theoretical AOI shift 0º-5º: | 1.55 | |||||||||||
| (following table represents filter data at 77K & AOI = 5 degrees, ALL RAW DATA in worksheets attached)) | ||||||||||||
| Position | center | edge1 | edge2 | edge3 | STATUS | |||||||
| T% | 83.20 | 82.59 | 83.02 | 82.76 | PASS | |||||||
| CWL(nm) | 2121.63 | 2121.45 | 2121.58 | 2121.85 | PASS | |||||||
| FWHM(nm) | 6.93 | 6.93 | 6.93 | 6.93 | PASS | |||||||