H2-2.12

Summary Part name:  2121.8/7 BARR ED086-1
Measurement: All parts are measured at 4 points. Once at Center and then at three points near the edge. In an f/8 beam at AOI=0
Measured 77K shift for T%:  0.418137
Measured 77K shift for CWL:  0.35
Measured 77K shift for FWHM:  0.1
Theoretical AOI shift 0º-5º: 1.55
(following table represents filter data at 77K & AOI = 5 degrees, ALL RAW DATA in worksheets attached))
Position center edge1 edge2 edge3 STATUS
T% 83.20 82.59 83.02 82.76 PASS
CWL(nm) 2121.63 2121.45 2121.58 2121.85 PASS
FWHM(nm) 6.93 6.93 6.93 6.93 PASS