FeII-1.65

Summary Part name:  1650.1/5.5 BARR ED090-1
Measurement: All parts are measured at 4 points. Once at Center and then at three points near the edge. In an f/8 beam at AOI=0
Measured 77K shift for T%:  -4.88558
Measured 77K shift for CWL:  0.15
Measured 77K shift for FWHM:  0.1
Theoretical AOI shift 0º-5º: 1.2
(following table represents filter data at 77K & AOI = 5 degrees, ALL RAW DATA in worksheets attached))
Position center edge1 edge2 edge3 STATUS
T% 81.93 81.08 81.14 81.27 PASS
CWL(nm) 1650.35 1650.15 1650.25 1649.95 PASS
FWHM(nm) 5.3 5.3 5.3 5.3 PASS