FeII-1.64

Summary Part name:  1644/5.5 BARR ED101-1
Measurement: All parts are measured at 4 points. Once at Center and then at three points near the edge. In an f/8 beam at AOI=0
Measured 77K shift for T%:  -3.04427
Measured 77K shift for CWL:  0.15
Measured 77K shift for FWHM:  0.1
Theoretical AOI shift 0º-5º: 1.19
(following table represents filter data at 77K & AOI = 5 degrees, ALL RAW DATA in worksheets attached))
Position center edge1 edge2 edge3 STATUS
T% 79.66 82.52 82.43 82.78 PASS spec: >60% (goal > 70%)
CWL(nm) 1644.16 1643.86 1643.96 1643.96 PASS spec: 1644nm +/- 0.3nm
FWHM(nm) 5.5 5.6 5.6 5.6 PASS spec: 5.5nm +/- 0.5nm