Summary |
|
Part
name: |
1644/5.5 BARR
ED101-1 |
|
|
|
|
|
Measurement: All parts are measured at 4 points. Once at Center and
then at three points near the edge. In an f/8 beam at AOI=0 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Measured 77K shift for T%: |
-3.04427 |
|
|
|
|
|
|
|
|
|
Measured 77K shift for CWL: |
0.15 |
|
|
|
|
|
|
|
|
|
Measured 77K shift for FWHM: |
0.1 |
|
|
|
|
|
|
|
|
|
Theoretical AOI
shift 0º-5º: |
1.19 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
(following
table represents filter data at 77K & AOI = 5 degrees, ALL RAW DATA in
worksheets attached)) |
|
|
|
Position |
center |
edge1 |
edge2 |
edge3 |
STATUS |
|
|
|
|
|
|
|
T% |
79.66 |
82.52 |
82.43 |
82.78 |
PASS |
|
spec: >60% (goal > 70%) |
|
|
|
CWL(nm) |
1644.16 |
1643.86 |
1643.96 |
1643.96 |
PASS |
|
spec: 1644nm +/- 0.3nm |
|
|
|
FWHM(nm) |
5.5 |
5.6 |
5.6 |
5.6 |
PASS |
|
spec: 5.5nm +/- 0.5nm |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|