| Summary | Part name: | 1644/5.5 BARR ED101-1 | ||||||||||
| Measurement: All parts are measured at 4 points. Once at Center and then at three points near the edge. In an f/8 beam at AOI=0 | ||||||||||||
| Measured 77K shift for T%: | -3.04427 | |||||||||||
| Measured 77K shift for CWL: | 0.15 | |||||||||||
| Measured 77K shift for FWHM: | 0.1 | |||||||||||
| Theoretical AOI shift 0º-5º: | 1.19 | |||||||||||
| (following table represents filter data at 77K & AOI = 5 degrees, ALL RAW DATA in worksheets attached)) | ||||||||||||
| Position | center | edge1 | edge2 | edge3 | STATUS | |||||||
| T% | 79.66 | 82.52 | 82.43 | 82.78 | PASS | spec: >60% (goal > 70%) | ||||||
| CWL(nm) | 1644.16 | 1643.86 | 1643.96 | 1643.96 | PASS | spec: 1644nm +/- 0.3nm | ||||||
| FWHM(nm) | 5.5 | 5.6 | 5.6 | 5.6 | PASS | spec: 5.5nm +/- 0.5nm | ||||||