| Summary | Part name: | 1599.5/5 BARR ED113-1 | |||||||||
| Measurement: All parts are measured at 4 points. Once at Center and then at three points near the edge. In an f/8 beam at AOI=0 | |||||||||||
| Measured 77K shift for T%: | -4.0879 | ||||||||||
| Measured 77K shift for CWL: | 0.3 | ||||||||||
| Measured 77K shift for FWHM: | 0.1 | ||||||||||
| Theoretical AOI shift 0º-5º: | 1.16 | ||||||||||
| (following table represents filter data at 77K & AOI = 5 degrees, ALL RAW DATA in worksheets attached)) | |||||||||||
| Position | center | edge1 | edge2 | edge3 | STATUS | Spec's | |||||
| T% | 79.80 | 80.36 | 79.84 | 80.50 | PASS | > 70% | |||||
| CWL(nm) | 1599.24 | 1599.64 | 1599.44 | 1599.24 | PASS | 1599.5+/-0.3 | |||||
| FWHM(nm) | 5.1 | 5 | 5 | 5.1 | PASS | 5+/-0.5 | |||||