BrG-2.17

Summary Part name:  2174.2/7 BARR ED087-1
Measurement: All parts are measured at 4 points. Once at Center and then at three points near the edge. In an f/8 beam at AOI=0
Measured 77K shift for T%:  -0.8669
Measured 77K shift for CWL:  0.3
Measured 77K shift for FWHM:  0
Theoretical AOI shift 0º-5º: 1.59
(following table represents filter data at 77K & AOI = 5 degrees, ALL RAW DATA in worksheets attached))
Position center edge1 edge2 edge3 STATUS
T% 87.56 84.98 85.42 85.24 PASS
CWL(nm) 2173.91 2174.31 2174.01 2173.91 PASS
FWHM(nm) 7.2 7.3 7.3 7.2 PASS