Summary |
|
Part
name: |
2166.1/7 BARR
ED107-1 |
|
|
|
|
|
Measurement:
All parts are measured at 4 points. Once at Center and then at three points
near the edge. In an f/8 beam at AOI=0 |
|
|
|
|
|
|
|
|
|
|
|
|
Measured 77K shift for T%: |
-0.22052 |
|
|
|
|
|
|
|
|
Measured 77K shift for CWL: |
0.35 |
|
|
|
|
|
|
|
|
Measured 77K shift for FWHM: |
0.1 |
|
|
|
|
|
|
|
|
Theoretical
AOI shift 0º-5º: |
1.58 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
(following
table represents filter data at 77K & AOI = 5 degrees, ALL RAW DATA in
worksheets attached)) |
|
|
Position |
center |
edge1 |
edge2 |
edge3 |
STATUS |
|
|
|
|
|
|
T% |
88.63919 |
88.61839 |
88.53367 |
88.67765 |
PASS |
|
|
|
|
|
|
CWL(nm) |
2166.35 |
2166.35 |
2166.15 |
2166.25 |
PASS |
|
|
|
|
|
|
FWHM(nm) |
6.9 |
6.9 |
7 |
6.9 |
PASS |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|